IVDR Amendments
On smaller screens, scroll horizontally to view all four columns.
| Topic | Current IVDR Requirements | Proposed Changes (Dec. 16, 2025) |
Impact/Intent |
|---|---|---|---|
| Notified Body (NB) Certification | CE certificates valid for max 5 years with full recertification. | No fixed expiry; replaced by risk-based periodic reviews. | Reduces re-certification burden and bottlenecks with Notified Body reviews. For companion diagnostics, this represents a significant improvement with co-development efforts. |
| NB Surveillance Audits | Typically annual onsite audits. | Less frequent / flexible, including remote audits. | Improves efficiency, reduces costs. |
| Technical Documentation Review | Regular NB review of representative devices (esp. Class B/C/D). | Risk-triggered reviews only (not routine). | Reduces burden associated with periodic reviews and allows the NB to focus effort on higher-risk concerns. |
| Performance Evaluation / Clinical Evidence | Strict performance study and clinical evidence requirements; limited use of alternatives. | Expanded acceptance of non-clinical data (e.g., modeling, simulations) and more flexibility for establishing clinical equivalence with other CE marked IVDs. | Reduces operational and cost burdens associated with prospective performance studies and enables faster IVD development and innovation. |
| Performance Studies (IVDs) | Authorization and notification by local competent authority required for many studies. | Exemptions for low-risk studies (e.g., routine blood draws) and simplified requirements for CDx studies using leftover samples. | Exemptions eliminate the need for competent authority review and authorization for certain low-risk performance studies; Streamlines clinical evidence generation; reduces burden associated with IVD/Drug sponsor co-development. |
| Companion Diagnostics (CDx) Performance Studies | Separate regulatory processes for CDx performance studies and corresponding drug trials. | Single application for combined studies (device + drug) with coordinated assessment by regulators. | Reduces administrative burden and cost; accelerates development timelines; improves alignment between drug and CDx approval. |
| Post-Market Surveillance (PMS) / PSUR | Frequent updates (e.g., annual for higher classes) and NB review obligations. | Reduced frequency (e.g., every 2 years) and less intensive NB review. | Reduces lifecycle regulatory burden. |
| In-house (Health Institution) IVDs | Strict conditions; generally no transfer outside institution. | More flexibility, including transfer in public health interest; removal of requirement that there is no CE marked equivalent on the market to justify use of in-house IVD. | Improves access and emergency preparedness; |
| Classification Rules | Conservative classification; many devices require NB involvement. | Adjusted classification rules, potentially lowering class for some products (e.g., software). | Down classification of IVDs reduces unnecessary NB involvement. |
| Conformity Assessment procedures for breakthrough devices or orphan devices | Limited flexibility for niche or innovative devices. | Criteria for breakthrough devices and orphan devices introduced; breakthrough devices and orphan devices will be subject to a priority and rolling review; manufacturers have access to expert panels’ advice. | Introducing expedited pathways for IVDs with a breakthrough device or orphan device designation will accelerate development and commercialization efforts for these IVD types. |
| Software / Digital Health | Often classified at higher risk (Class C/D). | More proportionate classification for software. | Supports digital innovation. |
| Artificial Intelligence (AI) | Devices may fall fully under EU AI Act requirements. | Reduced overlap with AI Act; limited applicability. | Avoids regulatory duplication. |
| Person Responsible for Regulatory Compliance (PRRC) | Strict qualification requirements and availability obligations. | Relaxed requirements, especially for SMEs. | Reduces burden on small companies and allows more flexibility with identifying PRRC candidates. |
| Use of Electronic IFU (eIFU) | Limited acceptance. | Expanded use of eIFU. | Supports digital transformation and allows for instant implementation of labeling changes in cases where eIFUs are allowed. Manufacturing costs are also lowered. |
| Structured Dialog with Notified Bodies | N/A | Requirement added for notified bodies to have documented procedures in place to offer and carry out dialog with the manufacturer before and after an application for conformity assessment. | The new requirement clarifies the need for notified bodies to have a formal process for structured dialog and for the notified bodies to offer structured dialog before and after applications for conformity assessment are filed. Prior to this, MDCG 2019-6 described limited scenarios where a sponsor could seek structured dialog with a notified body before an application is filed. |
| Advice from Expert Panels | N/A | For class C and class D devices, a sponsor may, prior to its performance study or clinical performance evaluation, consult an expert panel to solicit feedback on a study protocol prior to initiating the study. The guidance from the panel will be considered by both the sponsor and notified body when reviewing subsequent performance evaluation reports as part of the conformity assessment. | Introducing an expert panel review prior to initiating performance studies streamlines development efforts and avoids the potential for notified bodies to disagree with the study design during the conformity assessment review. |